Evaluation of Mycobacteria Growth Indicator Tube(MGIT), an Automated Culture System for Detection of Mycobacteria from Clinical Specimens.

Accession number;99A0385878
Title;Evaluation of Mycobacteria Growth Indicator Tube(MGIT), an Automated Culture System for Detection of Mycobacteria from Clinical Specimens.
Author; KOBAYASHI INTETSU (Mitsubishi Kagaku Bio-Clinical Lab., Inc.) TODA HARUYO (Mitsubishi Kagaku Bio-Clinical Lab., Inc.) KOYAMA ETSUKO (Mitsubishi Kagaku Bio-Clinical Lab., Inc.) HASEGAWA MIYUKI (Mitsubishi Kagaku Bio-Clinical Lab., Inc.) HASHIGUCHI NORISHIGE (Mitsubishi Kagaku Bio-Clinical Lab., Inc.) ARAI HIDEO (Fed. of Natl. Pubulic Serv. and Affil. Personnel Mutual Aid Assoc., Miyagino Hosp.) FUKAMI TOSHIE (Uenomiyakyokai Kiyoseuenomiyabyoin) WATANABE AKIRA (Res. Inst. Tuberculosis and Cancer, Tohoku Univ.)
Journal Title;Journal of the Japanese Association for Infectious Diseases
Journal Code:Z0760A
ISSN:0387-5911
VOL.73;NO.2;PAGE.172-178(1999)
Figure&Table&Reference;FIG.1, TBL.4, REF.20
Pub. Country;Japan
Language;Japanese
Abstract;We compared the Mycobacteria Growth Indicator Tube 960(MGIT 960) and Ogawa medium(OM) for the detection of mycobacteria (acid-fast bacteria: AFB) using 882 sputum specimens. Over-all, 120 strains of AFB were isolated by the MGIT 960 system and 99 strains of AFB were isolated by using OM. As far as Mycobacterium tuberculosis is concerned, 88 and 71 isolates were achieved by the MGIT 960 and OM respectively. A total of 28 isolates (18 isolates of M. tuberculosis and 10 isolates of nontuberculous mycobacteria: NTM) were detected by the MGIT 960 only whereas only 2 isolates (1 M. tuberculosis and 1 NTM) were detected by OM only. Of these sputum specimens, 72 were smear positive for AFB. The rates of smear negative but culture positive specimens were 8.0% (65 out of 809) for the MGIT 960 system and 6.2% (50 out of 809) for OM. The contamination rate for MGIT 960 was only 1.2%. The average time required for detection of M. tuberculosis was 14.1 days by the MGIT 960 system and 24.6 days by OM. For the NTM, the average detection time were 8.3 days for the MGIT 960 system and 22.8 days for OM. These results indicate that the MGIT 960 system allows detection of mycobacteria significantly faster than OM. (author abst.)
FULLTEXT