Accession number;02A0151823
|
| Title;Validation of the Accuracy of the LabSOCS Software for Mathematical Efficiency Calibration of Ge Detectors for Typical Laboratory Samples. |
| Author;
BRONSON F L
(Canberra Ind., 800 Res. Parkway, Usa)
|
Journal Title;Abstracts of Papers. Symposium on Radiochemistry
|
Journal Code:F0139B
|
ISSN:1345-2762
|
|
VOL.45th;NO.;PAGE.45(2001)
|
| Figure&Table&Reference; |
| Pub. Country;Japan |
| Language;English |
| Abstract;The LabSOCS software performs mathematical efficiency calibration of Ge detectors. Extensive tests have been conducted comparing the LabSOCS efficiency calibration with those from traditional radioactive sources for typical Laboratory geometries. The results from the first 13 detectors produced using this process show that efficiency calibrations generated with LabSOCS have an accuracy of 5.1% sd at low energies and 4.2% sd at high energies. (author abst.) |
|
|
|