Title;Development of a fluorescence detector equipped with Xe lamp and photon counting head for capillary electrophoresis
Author; NAKAJIMA RIICHIRO (Doshisha Univ., Fac. of Eng.) TSURUGI AKITSUGU (Doshisha Univ., Fac. of Eng.) NODA KEIICHI (Doshisha Univ., Fac. of Eng.) TANAKA TAKAYOSHI (Doshisha Univ., Fac. of Eng.) TSUKAGOSHI KAZUHIKO (Doshisha Univ., Fac. of Eng.)
Journal Title;Nippon Kagakkai Koen Yokoshu
Journal Code:S0493A
ISSN:0285-7626
VOL.84th;NO.1;PAGE.12(2004)
Figure&Table&Reference;FIG.1, REF.1
Pub. Country;Japan
Language;Japanese
Abstract;We have used RF-535 (Shimadzu) for HPLC as the fluorescence detector modified for capillary electrophoresis to detect a fluorescence labeled compounds. In this report, for high sensitive detector with select of wavelength, we developed the fluorescence detector for capillary electrophoresis using photon counting head, Xe lamp as a light source and band pass filter as a select of wavelength. As a result, sensitivity was decreased on the detector. However, we report a few knowledges on development of the detector. (author abst.)